6. Operating procedures
We’ll show the operating procedures for the instrument on each key on the front panel.
Warning Operation with each key should be made after interrupting the
measurement. If it is made during measurement, it may cause an
unexpected malfunction.
Key operation in Measurement mode can be effective by pressing it
for about one second or so.
61. A/ZA key
611. Operations in Measurement mode
At the same time the Tare weight cancellation (A/Z) function activates and load display value
changes into the net weight display value, A/Z” on the status display will light up and load
display will become “0”.
612. Operations in another modes
(1) Carrying set value
By pressing the A/Z A key with the condition of displaying various kinds of set values, the
flashing digits for set value will carry from 10
0
,10
1
,10
2
,10
3
to 10
4
in order.
(However, the range of carry differs depending on the number of digits of set values and yes/no
of polarity sign.)
(2) Change of function and so on
Executes change of various kinds of functions in Function mode, and also changes of function
during Check mode.
During the operation in the paragraph 612, the key is pressed for
more than 2 s, each key operation will be executed continuously at a
constant interval without the operation of ON/OFF.
62. A/Z OFF Y key
621. Operation in Measurement mode
Only when the Tare weight cancellation (A/Z) function is operating, tare weight cancellation
clear (A/Z OFF) operates and at the same time when load display value becomes gross weight
display value, A/Z” on status display will put off.