5−2−2. Calibration method to register the output of strain gage applied transducer at the time
of zero and the maximum display
Warning ● Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration.
If calibration shall not be made, correct measured results may not be
obtained nor may cause malfunction in the instrument and there
may exist damage in peripheral equipments.
Besides, even though calibration had been made, there may occur the
similar case when the result is not correct, so make precise calibration
again.
● During the calibration is executing, be sure to set Tare weight
cancellation clear, and to make cancellation (Execution of F−98) for
compensated data on zero set and also set the OFF position of Zero
tracking(Setting “00000” on F−07 and F−08 each).
If neglected, correct measured results may not be obtained.
● During calibration procedures, press the “FUNC.” key to interrupt
the calibration. It will hold the calibration data before interrupting
and will return to the Measurement mode.
● When the “A/Z A ” key is pressed with the load display of “FUNC.”,
the display will change as the following sequence at every time the
key is pressed. “FUNC”→“CCAL”→“A CA L ” →“LCAL”→“ZERO”→
“SPAN”→“TARE”→“CHECK”→“MONIT”→“F−END”→“FUNC”→
“CCAL”→・・・・・(Hereinafter, it will repeat.)