73
Procedures
1
Press the “FUNC.” key for about one second.
The load display will show “FUNC”.
2
Press the A/Z A key at 8 times.
The load display will change as “CCAL
“A CA L “LCAL“ZERO”“SPAN”“TARE”
“CHECK”“MONIT”.
3
Press the “ENTER key.
Monitor mode can be entered, and converted
value into mV/V for the present input value for
the transducer will flashing display on the load
display.
4
When quitting the Monitor mode, press the
“ENTER key. The Measurement mode can be
returned and load value will be shown.