73
Procedures
1
Press the “FUNC.” key for about one second.
The load display will show “FUNC”.
2
Press the “A/Z A” key at 8 times.
The load display will change as “CCAL ”→
“A CA L ” →“LCAL”→“ZERO”→“SPAN”→“TARE”
→“CHECK”→“MONIT”.
3
Press the “ENTER” key.
Monitor mode can be entered, and converted
value into mV/V for the present input value for
the transducer will flashing display on the load
display.
4
When quitting the Monitor mode, press the
“ENTER” key. The Measurement mode can be
returned and load value will be shown.