29
2 Basic Measuring Methods
No. 99MAF029B
2.5
Measuring the Plane Displacement (Plane
Scanning Measurement)
Moving the probe while maintaining contact with the surface of the workpiece obtains multiple
measurements. Based on the acquired data, the maximum, minimum, and displacement values
(maximum – minimum = displacement) can be measured.
As an example, the procedure for measuring the displacement of a1 (maximum value), a2 (minimum
value), and A (1 mm), shown below, is explained.
A (1 mm)
a2
a1
ABS origin
NOTE
Measure within a range of about 1 mm from the starting point. Measuring with a wider range may
increase measurement errors.
1 mm
1
Press .
»
will be displayed on the upper left of the LCD screen.
2
Begin measuring.
1
Slowly bring the probe into contact with
the measuring surface until a beep
sound is made.
Maintain contact between the probe and
the surface.
»
[0.000] will be displayed on the
LCD screen.
2
Secure the probe up/down wheel with
your hand or the clamp screw in order to
prevent it from moving.

Table of Contents

Other manuals for Mitutoyo QM-Height Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mitutoyo QM-Height Series and is the answer not in the manual?

Mitutoyo QM-Height Series Specifications

General IconGeneral
BrandMitutoyo
ModelQM-Height Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals