Verify that all of the bars on the LCD are at or below the horizontalline.
Record result on the TDR.
6.6
Current Consumption Test
6.6.1
Leave TF-0346 connected to the UUT.
6.6.2
Connect DMM to TF-0347 and setup meter to read current.
6.6.3
Set the current switch on TF-0347 to measure.
6.6.4
Exit lead quality screen on UUT and enter the display ECGscreen.
6.6.5
Verify the current is below 85mA.
6.6.6
Record the current draw on the TDR.
6.7
Recording Test
6.7.1
Remove TF-0347 and TF-0346 from UUT.
6.7.2
Connect the UUT to the simulator.
6.7.3
Put a new AA battery into the UUT and install the batterycompartment.
6.7.4
Select the record function on the UUT.
6.7.5
Allow the UUT and simulator to run for approximately 2 minutes.
6.7.6
Remove the compact flash card and review on an H-Scribe system. Verify the UUT
properly recorded for the approximate 2 minutes. Record a pass or failure on the
TDR.
6.8
Clock Time Retention
6.8.1
Verify the time and date on the UUT are correct.
6.8.2
If incorrect, set them to the correct value and exit the menu option.
6.8.3
Cycle the UUT power and recheck to verify the correct date and time were retained.
6.8.4
Record the result on the TDR.
6.9
LCD Test
6.9.1
Verify that throughout the test no missing or extra segments on the UUT’s LCD were
observed. Record a pass or failure on the TDR.