10 Maintenance
Messko GmbH 202070 BA4001150/10 ENMSENSE
®
DGA 2/3
10.3 Oil extraction
Figure45: Oil extraction
The scope of delivery includes two MESSKO® sample extraction kits for tak-
ing oil samples via the oil sample extraction connection. For information on
taking oil samples, refer to the included instructions BA2938054 and chapter
Device field calibration [►Section 10.4, Page 71] in these instructions.
Send the oil sample, together with the completed "SAMPLE DATA SHEET –
OIL ANALYSIS", to MESSKO GmbH for oil analysis. If the results of the
analysis indicate that the device requires a field calibration, you will receive
an action recommendation for the settings from MESSKO through the
MESSKO® MSET parameterization software.