EasyManua.ls Logo

Myron L 750 II Series - Optional Features; Accessories

Myron L 750 II Series
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NOTE: 3 1/2 digit LCD may be replaced with either a 3 1/2 or a
4 1/2 digit backlit LCD - available as options.
ACCURACY
Analog Models: + 2% of Full Scale
Digital Models: + 1 % of Full Scale
SENSITIVITY
0.05% of span
STABILITY
0.05% of span
REPEATABILITY
0.1% of span
CALIBRATION CHECK
Built in full scale value
RECORDER OUTPUT
0-10VDC@5mAmax.(linear);standardonallmodels
OUTPUT IMPEDANCE
100Ω±5%
SENSOR INPUT 1 (optional 3 sensor input available)
CONTROL FUNCTION
Models 757II, 758II & 752II, 753II:
Single set point alarm/control continuously adjustable
0-100% of span
Second set point alarm/control (Optional).
Continuously adjustable - 0-100% of span
Hysteresis
Adjustable from 0.3-3% of full scale
Indicators
“HIGH” (red) and “LOW” (green) set point LEDs - reversible.
Relay Contact Rating
SPDT10amp@250VAC,30VDC.Relayoperates
increasing or decreasing reading (user selectable).
Solid State Output
24VDCunregulated,30mAMaximum.
Powers optional PA - Piezo Electric Alarm, RA™ - Remote
Alarm, or ?
SecondAlarm/controlModule,withabovespecications,opt.on
Models 757II, 758II & 752II, 753II
POWER SPECIFICATIONS
115/230 VAC±15%,50/60Hz(Userselectable)
overvoltage category II
100 mA Maximum Current
Double Insulated (with circuit board ground for OEM
operation)
Fuse-100mASlowBlow(T.10A)forboth115&230VAC
(V~)Userreplaceable
Humidity - 20-90% non-condensing
Max. Altitude -
12,000 meters/40,000 ft non-operating
3000 meters/10,000 ft operating
Pollution degree 2
24 VAC or 24 VDC Option available on Special Order
OvervoltagecategoryII(24VAC)
250 mA Maximum Current
Double Insulated (with circuit board ground for OEM
operation)
Fuse - 250mA Slow Blow (T.25A) User replaceable
AMBIENT TEMPERATURE RANGE
0-60ºC/32-140ºF
DIMENSIONS
152H x 122W x 96 D mm / 6.0 H x 4.8 W x 3.8 D in.
HOUSING CONSTRUCTION
Fully gasketed heavy-duty ABS for splash-proof and corrosion
resistance.
Rated IP65/NEMA 4X
Double Insulated (with circuit board ground for OEM operation)
WEIGHT
750 Series II: average 0.9 kg/2 Ibs.
C. OPTIONAL FEATURES
SeesectionsVI&VIIforcompleteselectionanddetails.
-SC Second Alarm/control Module (M/c only)
-4A 4-20 mA Isolated output (digital models only)
-PA 70 db Piezo Electric Alarm (digital M/c only)
-PAT Piezo Alarm & Timer Module (digital M/c only)
-45BL 4 1/2 Digit Backlit Liquid Crystal Display (LCD)
-35BL 3 1/2 Digit Backlit Liquid Crystal Display (LCD)
-TP Temperature Module 0-200ºC, requires -TP sensor
(digital Monitor/controllers only)
-TH Alarm/control Interface Harness (req. SC & TP)
(digital only)
-PC 115VACPowercord(8ft./2,44meters)withUSA
plug and strain relief - NOTforusewith230VAC.
-24VA 24VACIsolatedPowerSupply(specialorder)
-24VD 24VDCIsolatedPowerSupply(specialorder)
-PTS Panel mounted Full Scale Test switch (special order)
D. ACCESSORIES (ordered separately)
SeesectionsVI&VIIforcompleteselectionanddetails.
ASI AQUASWITCH I, auto RO/DI switching system (req. M/c)
SCO Second Alarm/control Module (Monitor/controllers only)
4AO 4-20 mA Self-Powered Isolated Output Module
(digital only)
3SE 3 Sensor Input Module with Enclosure.
3SO 3 Sensor Input Module with 1 meter/3 foot cable*.
3RE 3 Range Switch with Enclosure†
3RO 3 Range Switch with 1 meter/3 foot cable†*.
3SRO Above 3SO & 3RO combined with enclosure†.
3SRE Above 3SO & 3RO combined†*.
PAO 70 db Piezo Electric Alarm only (753II/758II)
PATO Piezo Alarm & Timer Module (753II/758II)
45BLO 4 1/2 Digit Backlit Liquid Crystal Display (LCD)
35BLO 3 1/2 Digit Backlit Liquid Crystal Display (LCD)
TPO Temperature Module 0-200ºC - requires -TP
sensor (digital models only)
THO TP/TPO Alarm/control Harness only
(requires SCO & TPO)
PCO PowerCord115VAC(8ft./2,44meters)withUSA
plug and strain relief - NOTforusewith230VAC.
SMP50 Surface Mounting Plate
RA Remote Alarm™ - RA (Monitor/controllers only)
VR PowerSupply,24VAC,20VA.(115VAC,indooruse)
024-1 1in.NPT24VACSolenoidValve
CS-11 20ΩResistivityCalibrationModule(NISTTraceable)
CS-14 2ΩResistivityCalibrationModule(NISTTraceable)
CS-17 200ΩResistivityCalibrationModule(NISTTraceable)
†Add C for Conductivity or R for Resistivity to part number
*Customer mounted separately
6
DOUBLE INSULATED

Table of Contents

Related product manuals