3. Contraindications and Warnings
3.1. Contraindications
Natus Quantum systems shall NOT be used in the following conditions:
Do NOT operate the system in the presence of flammable anesthetics.
Check areas of use to avoid using the system in the presence of flammable gases.
To ensure the validity of signals, do not operate the device near any sources of
electromagnetic interference.
Natus systems are not AP or APG rated. DO NOT USE a Natus system in the presence of
a flammable anesthetic mixture with air, oxygen, or nitrous oxide.
The Quantum amplifier system is NOT designed to work with defibrillators. The system
could be damaged when used with this device.
Operation of this equipment with input signals in excess of a range of +/-10mV may cause
incorrect results.
3.2. Warnings and Cautions
3.2.1. General Warnings
NOTE: It is recommended that all data be stored using redundant storage capabilities. This
can help to minimize data loss in the event of a failure of the primary drive.
The equipment/system is intended for use by trained users. Please read the manual before
installing any of the hardware, and refer to the appropriate section when you operate, store
or re-install the system.
Only use the Quantum amplifier system in conjunction with approved devices and
accessories.
Use of cables other than those specified or sold by the manufacturer on the equipment,
may result in increased emissions or decreased immunity of the equipment and may cause
the system to be non-compliant with the requirements of IEC 60601-1-2:2007.