Natus Quantum Amplifier User & Service Manual
46
7.6. Quantum Amplifier System Connection Diagram
NOTE: Setup and Installation of the Quantum Amplifier System should be performed by Natus
qualified personnel only.
NOTE: When the Base Unit is connected to an acquisition computer via USB, the maximum
sample rate is 2048Hz. When the Base Unit is connected to an acquisition computer via
gigabit Ethernet, sample rates up to 16KHz may be used for 128 and 256 channel
configurations.
For systems used in the Operating Room with a Cautery device, the use of Ethernet type
connection is mandatory to ensure continuous recording.