Natus Quantum Amplifier User & Service Manual
20
Using the Quantum amplifier system with cables and accessories not approved by Natus
may negatively affect EMC performance, including electromagnetic immunity. Refer to the
Electromagnetic Immunity table in the Safety & Standards Conformity section.
External equipment may interfere with the performance of the Quantum amplifier system,
even if other equipment complies with CISPR Emission requirements. Refer to the EMC
Standard of Compliance and Normative References table in the Safety & Standards
Conformity section.
The Quantum amplifier system should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the amplifier should be observed to
verify normal operation in the configuration in which it will be used. Refer to the
Recommended Separation Distances table in the Safety & Standards Conformity section for
minimum recommended separation distances.
Do not connect items which are not specified as part of the Quantum amplifier system to the
system.
3.2.3. Electrical Warnings and Cautions
Natus systems are intended for connection to a properly grounded electrical outlet only.
Conductive parts of electrodes and their connectors are not to contact other conductive parts
and earth.
Do not place MULTIPLE PORTABLE SOCKET-OUTLETS (MPSOs) on the floor.
Do not connect additional MPSOs or extension cords to the Quantum Amplifier system.
ELECTRICAL SHOCK HAZARD: Do NOT turn on the system power until all cables have
been connected, verified and visually inspected for any damage. Failure to inspect the cables
may result in electrocution. Verification of electrical safety should be performed routinely.
ELECTRICAL SHOCK HAZARD: Do NOT service the system. Refer servicing to qualified
personnel only. Do NOT use repaired components without proper testing.
Do not use the MPSO with the Quantum Amplifier system for supplying power to any
equipment that is not part of the system.