Parameter Default Condition
D-BUF-SIZE 100
Data Store Units Same as UNITS
EXT OFF
EXT TRlG IN
TRIGGER OUT
FALLING
TRIG ON CMPLT
TRIG OUT POL TRIG ACTIVE LO
TRIG OUT FREQ FREQ
=
30 Hz
TRIG LEVEL
BAR GRAPH
LVL
=
0.001
W
OFF
GPlB ADDR 05
BAUD RATE 9600
These parameters adopt the following default values at power up and are not
affected by recalling a configuration or by setting the configuration to default.
6232 ECHO OFF
Local Lockout OFF
Display Brightness NORMAL
Table 9. below lists the menu commands effecting the saving and recalling of
meter configurations.
Table
9
-Meter Configuration Operations
Menu Operation Keypad Commands
Associated Remote Commands
SAVE CONFIG
Adjust SAVE TO
d,
*SAV
and ENTER
RECALL CONFIG Adjust RECALL
d,
*RCL,*RST
and ENTER
2.4.4
AUTO
CAL
The AUTO CAL command causes the 2832-C to perform AID conversions of
amplifier offset voltages (zero errors) arising from aging and temperature
effects. These conversions are then used in subtracting the appropriate error
voltage from each reading during normal operation. The 2832€ automatically
performs this procedure every time it powers up (or
is
reset). To achieve
stable reading at the specified accuracy, AUTO CAL should be executed for
each channel after a minimum 60 minute warm-up period from power-up.
Simply press ENTER when AUTO CAL
is
displayed. The display message area
will display "ONE MOMENT", followed by a buzzer sound indicating that
AUTO CAL is complete. Low Power detectors do not have to be connected to
the 28324 or removed from the radiation source to effectively execute AUTO
C AL.