EasyManuals Logo

Newport 2832-C User Manual

Newport 2832-C
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
Device Independent Commands
'CAL?
Calibration Query
Syntax:
*CAL?
Parameters:
None
Function:
This routine performs an auto calibration sequence on both channels. The
calibration sequence measures the amplifier
offsets
at all gain settings. These
measurements are stored and subtracted from subsequent measurements.
Acquisition is suspended during auto calibration and no other remote commands
will
be
processed until the auto calibration sequence is complete.
Returns: <number>
<number> represents the integer
0.
This return value is generated when
autocalibration
is
complete.
Related Commands: None
'CIS
Clear Status
Syntax:
*CIS
Parameters: None
Function:
This command clears the Standard Event Status register, device event register, and
the error queue.
As
a result
of
this command the status byte
is
cleared except for
the message available bit @it
4).
Because this command does not clear the input
and output buffers the MAV bit will not
be
affected. The *CLS command also
cancels any outstanding *OPC and OPC? commands.
Related Commands:
*ESR?, EVENT?, *IST?,
*
STB?
'ERR?
Error Query
Syntax:
*ERR?
Parameters:
None
Function:
Returns (and removes) oldest error message from the error queue. The error
message
is
made up of an error code and a text description
of
the error
(see
the Appendix B). A maximum of
10
errors can be stored in the queue.
If
no
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com