'RCL
Recall
Meter Configuration
Syntax:
*RCL
<config>
<config> is of type <number> which rounds to an integer in the range
0
to
9.
This number specifies a configuration storage buffer. The number
0
stands for
the default configuration buffer.
Function:
This command configures the meter to a configuration previously stored in
non-volatile memory (see *SAV), or to the default configuration for the
attached detector. This command will fail
if
the requested configuration
is
incompatible with the detector attached to the meter or empty. The param-
eters affected by this command are those listed in Table
8.
Returns: None
Related Commands: *RST, *SAV
'ST
Reset
Syntax:
*RST
Parametem
None
Function:
This sets the meter to the default configuration,
see
Table
8,
for the detector
attached to it. It also cancels any outstanding *OPC or *OPC? commands.
Related Commands:
*RCL, *SAV
'
SAV
Save Meter Configuration
Syntax:
*SAV
<config>
Parametels:
<config>
is
of type <number> which rounds to an integer in the range
1
to
9.
This number specifies a configuration storage buffer.
Function:
This command saves the configuration of the meter to the configuration
specified by the parameter. The parameters saved by this command are
those listed in Table
8.
Related Commands: *RCL, *RST