DSCNT-n?
Data Store Value Count Query
Syntax:
DSCNT-n? <number>
where n is either
A
or
B
Parameters:
None
Function:
This query returns the number of data values stored in the data store buffer.
Returns:
<count>
<count>
is
of type <number> and represents an integer in the range of
1
to the
size of the data store buffer.
Related Commands:
DS-n?, DSCLR-n, DSE-n, DSE-n?, DSSIZE-n, DSSIZE-n?,
DSUNITS-n?, DSBUF-n, DSBUF-n?
DSE-n
Data Store Enable
Syntax:
DSE-n <state>
where n
is
either
A
or
B
Parameters:
The parameter <state> is a <number> which is either
0
or
1.
If
the number is
0,
data storing
is
disabled.
If
the number
is
1,
data storing is enabled.
Function:
This command enables or disables data storing. Note that data storing cannot
be enabled when the data store buffer
is
full and FIXED BUFFER behavior is
enabled. The user must first either clear the data store buffer or enable SLIDE
BUFFER behavior and then enable data storing.
Related Commands:
DS-n?, DSCLR-n, DSCNT-n?, DSE-n?, DSSIZE-n,
DSSIZE-n?, DSUNITS-n?, DSBUF-n, DSBUF-n? DS-n
DSE-n?
Data Store Enable Query
Syntax:
DSE-n?
where n is either
A
or
B
Parameters:
None
Function:
This query returns a value showing whether or not data storing is enabled or
disabled.
Returns:
<state>
<state>
is
of type <number> and represents an integer
0,
if data storing is
disabled, or
1,
if
data storing
is
enabled.
Related Commands:
DS-n?, DSCLR-n, DSCNT-n?, DSE-n, DSSIZE-n,
DSSIZE-n?, DSUNITS-n?, DSBUF-n, DSBUF-n? DS-n
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com