REFSEL-n?
Reference
Source
Query
Syntax: REFSEL-n?
Parameters: None
Function:
This query returns the current reference source that will be used in linear
and
logarithmic (dB) relative measurements. Returns:
<source>
is
of type <string> in the double quote format. The possible values
of <source> are:
"STOREF"
-
Last stored reference measurement
"OTHERCH"
-
The other channel
"USRREF"
-
User defined reference value
Related Commands: REFSEL-n, STOREF-n, USRREF-n, USRREFn?
RESP-n?
Responsivity
Query
Syntax: RESP-n?
where n is either
A
or
B
Parameters: None
Function:
This query returns the responsivity value currently being used by the
2832-C
in measurement calculation.
Returns: <val>
<val>
is
of type <number> and represents a floating point number. The units
of the responsivity <val> depends upon the type of detector family of the
detector in use. See Table
22
for a listing of responsivity units by detector
family.
Related Commands:
ATN-n, ATN-ri?, LAMBDA-n, LAMBDA-n?, USRCALn,
USRCAL-n?, USRRESP-n, USRRESP-n?
RUN
Start Both Channel Acquisition
Syntax: RUN
Parameters:
None
Function:
This command initiates data acquisition on both channels using the currently
defined acquisition mode. In single type acquisition modes this command will
cause one reading to be acquired. In continuous type acquisition modes this
command will initiate the continuous acquisition of readings. The STOP
command will terminate acquisition.
Related Commands: STOP, STOP-n, RUN-n