The display shows:
"CAL MODULE
A"
Make sure the CAL MODULE is seated
or "CAL MODULE B" correctly and reset the meter.
Try
another
and "MISSING OR BAD" CAL MODULE, if one is available, to
localize the problem to the first
CAL MODULE.
If
the first CAL MODULE does not work
after proper seating and reset, but the
second CAL MODULE works, call Newport
Customer Service and arrange for a repair
of the first CAL MODULE.
If
neither the first or second CAL MODULE
works, call Newport Customer Service
and arrange for a meter repair.
Be
sure to
also return the first CAL MODULE
if
you
cannot otherwise determine that it
is
working.
The display shows:
"CAL MODULE
A
"CAL MODULE B"
and "CHECKSUM ERROR
The display shows:
"BATTERY FAILURE"
RS232
communication does
not seem to work
No listener error when attempting
GPlB communications
Display value does not change.
Reading is different than
expected.
Call Newport Customer Service and or
arrange to return the meter for repair.
Call
Newport Customer Service and
arrange to return the meter for battery
replacement.
Check the
RS232
cable connection and
cable pinouts, Figure
10.
Make sure that
the device talking to the meter
is
setup for
8
data bits, no parity, and
1
stop bit. Also
check the echo mode state.
Check the GPlB cable connections and
the GPIB address.
Press the R/S, R/S A, or R/S B key.
See Table
19,
Common Measurement
Errors. (Section
3.10.6.)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com