IMPORTANT NOTES
1.7. Check scope of delivery
Open the packaging with care. Avoid fingerprints and sweat on the camera lens. Pre-
vent the camera and some accessories from falling and being damaged.
Remove all components from the packaging and check for completeness according to
the delivery note. The microscope could be packaged in more than one carton depend-
ing on the model or accessories.
Handle the microscope with care and always protect it against violent impact and
vibrations.
Keep the original packaging for possible longer storage or return of the device to the
supplier in case a repair or revision is needed.
Never lift the microscope by the illumination unit, cross stage, viewing head or
coarse/fine focusing knobs. This can damage the microscope!
Always carry the microscope with both hands.
Only transport the microscopes of the NIB600 series using the carrying handles
provided.
1.8. Site location
Select a suitable location before setting up the microscope.
Place the microscope on a resistant surface away from strong sunlight, heat sources,
high humidity, high dust exposure as well as strong vibration.
Microscope usage environmental requirements:
Do not expose the device to temperatures below +10°C or higher than +35°C! If
the device is exposed to temperatures outside this range, this will cause irreparable
damage to optical and/or mechanical parts which are not covered by warranty.
Maximum relative humidity: 80% at 31°C, linear temperature down to 34°C is
70%, 60% at 37°C, 50% at 40°C.
Place the microscope on a surface that meets the following criteria:
Flat
Vibration-free
Level
Hard surface and non-flammable
Chemical and mechanical resistant
To avoid blocking the natural convection air for cooling, ensure at least 10 centimetres
of space around the microscope, from the wall and other objects.
Place the microscope so that you can disconnect it from the mains at any time.
The mains cable must be freely accessible at all times, as the mains cable is intended
as a disconnecting device from the mains.
Make sure any residual moisture is fully eliminated before use.
For use in warm and humid climates, all optical components of the microscope are
already equipped with anti fungus protection.
Grounding the microscope to avoid lightning strike