Integrator's Guide14
© 2017 Nortek AS
SETPLAN,MIAVG=600,AVG=1,DIAVG=0,VD=0,MV=10,SA=35,BURST=0,
MIBURST=120,DIBURST=0,SV=0,FN="Data.ad2cp",SO=0,FREQ=75
OK
SETAVG,NC=109,CS=5,BD=2,CY="ENU",PL=-6,AI=180,VR=1,DF=3,NPING=137,
NB=3,CH=0,MUX=0,BW="BROAD",ALTI=0,BT=0,ICE=0
OK
%Setting plan for "Coarse" profile
SETPLAN1,MIAVG=1800,AVG=1,DIAVG=0,VD=0,MV=10,SA=35,BURST=0,
MIBURST=120,DIBURST=0,SV=0,FN="Data.ad2cp",SO=0,FREQ=55
OK
SETAVG1,NC=54,CS=20,BD=2,CY="ENU",PL=-2,AI=180,VR=1,DF=3,NPING=60,
NB=3,CH=0,MUX=1,BW="NARROW",ALTI=0,BT=0,ICE=0
OK
%Setting the alternating measurement intervals and ratios of "Fine"
and "Coarse"
SETALTERNATE,EN=1,PLAN=1380,IDLE=10,PLAN1=180,IDLE1=230
OK
%Setting the telemetry file to average the "Fine" profile over the
averaging interval
SETTMAVG,EN=1,CD=1,PD=1,AVG=180,TV=1,TA=1,TC=1,CY="ENU",FO=1,SO=0,
DF=100
OK
SAVE,ALL
ERROR
%Finding where the error in the configuration is
GETERROR
"Invalid setting: Avg Average Interval too low for the configured
number of pings and profiling distance",LIM="GETAVG1LIM,AI=
([360;1800])"
OK
%Number of pings too high compared to desired averaging interval
with multiplex enabled.
SETAVG1,NPING=30
OK
SAVE,ALL
OK
Note that SETTMAVG,AVG must equal the AI set by SETAVG,AI. To set telemetry averaging for the
alternate plan use SETTMAVG1, note that these will be recorded to the same telemetryfile.bin
file.
Enter START or DEPLOY,TIME to begin the deployment.