5.12 GETMEM
Returns the amount of memory that will be stored on the recorder for the burst and average data
as well as the combined plan value. Alternate mode values is also supported
Argument
Description
PLAN
Combined burst and average memory [Mbytes/hour]
BURST
Burst memory [Mbytes/hour]
AVG
Average memory [Mbytes/hour]
PLAN1
Combined burst1 and average1 memory [Mbytes/hour]
BURST1
Alternate burst1 memory [Mbytes/hour]
AVG1
Alternate average1 memory [Mbytes/hour]
TOTAL
Total memory value [Mbytes/hour]
5.13 SETTRIG/GETTRIG/GETTRIGLIM
Sets / gets the parameters and limits for Trigger. The available trigger types will depend on the
harness/cable used with the instrument.
When triggered the instrument will perform a complete ping (Tx and Rx) before it goes back to
monitoring the trigger. Any triggers asserted during an ongoing ping will be ignored. There are no
specific requirements for pulse length. The unit triggers on the edge(s) of the trigger signal and
can be triggered on rising, falling or both edges.
Argument
Description
EN
Enable/disable trigger functionality
TRIG
Specifies trigger type
TRIGOUTPUT
Enable trigger output
Note: TRIGOUTPUT=1 enables master trig output, RS485EDGE trigger must be used with this
option. This enables several instruments to be synchronized together through RS485 with one of
the instruments acting as master. Only continuous measurement configurations are supported in
this mode.
5.13.1 TRIG
Command used for trigging measurement when Trigger is enabled and trigger type equals
“COMMAND”.
Argument
Description
ID
Counting number

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nortek Signature Series and is the answer not in the manual?

Nortek Signature Series Specifications

General IconGeneral
TypeAcoustic Doppler Current Profiler (ADCP)
Operating Temperature-5°C to +45°C
Storage Temperature-20°C to +60°C
Temperature SensorIntegrated
Pressure SensorIntegrated
Compass and Tilt SensorIntegrated
Data StorageInternal memory
Communication InterfacesEthernet, RS-232
Depth RatingUp to 6000 meters (depending on model)
Data OutputASCII, binary
Beam AngleVaries by model
Velocity Accuracy±0.5% of measured velocity ±1 mm/s