4.4 Erase telemetry file
In this example, the user wishes to erase the telemetry file after some period of time.
Send Break
CONFIRM
OK
%Going into Data Retrieval Mode
RM
NORTEK AS.
Version 2176 (Sep 17 2015 18:58:53)
DATA RETRIEVAL MODE
OK
%Checking the size of the telemetry file. Return in bytes
TMSTAT
34768
OK
%Erasing the telemetry file
ERASETM,9999
OK
%Continuing the configured deployment plan
CO
OK
The telemetry file can also be erased over FTP.