The INQ command inquires the instrument state. Note that when operating over RS232 or RS422
serial lines, it should be preceded with @@@@@@ <delay 400 ms> and a flush of the input
buffer in case the instrument is in power down or in a low power mode taking measurements.
Consult this section a description of the Instrument modes.
Parameter
Instrument Mode
0000
Bootloader/Firmware upgrade
0001
Measurement
0002
Command
0004
Data Retrieval
0005
Confirmation
0006
FTP-mode
Example (in command mode) :
08:43:31 INQ<CR><LF>
08:43:31 0002<CR><LF>
Example (in measurement mode) :
08:43:31 INQ<CR><LF>
08:43:31 0001<CR><LF>
Example (in confirmation mode) :
08:43:31 INQ<CR><LF>
08:43:31 0005<CR><LF>
Example (in data retrieval mode) :
08:43:31 INQ<CR><LF>
08:43:31 0004<CR><LF>
Example (in firmware upgrade mode) :
08:43:31 INQ<CR><LF>
08:43:31 0000<CR><LF>

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nortek Signature Series and is the answer not in the manual?

Nortek Signature Series Specifications

General IconGeneral
TypeAcoustic Doppler Current Profiler (ADCP)
Operating Temperature-5°C to +45°C
Storage Temperature-20°C to +60°C
Temperature SensorIntegrated
Pressure SensorIntegrated
Compass and Tilt SensorIntegrated
Data StorageInternal memory
Communication InterfacesEthernet, RS-232
Depth RatingUp to 6000 meters (depending on model)
Data OutputASCII, binary
Beam AngleVaries by model
Velocity Accuracy±0.5% of measured velocity ±1 mm/s

Related product manuals