5.31 GETSTATE
Returns information about the current operational state of the instrument
Argument
Description
MODE
Current instrument state
0001Measurement (START command received)
0002Command
0003Deploy (DEPLOY command received and deployment running)
0004Data retrieval
0005Confirmation
0006Network FTP
0008Pre-deployment (DEPLOY command received, but deployment has not
yet started)
0009 –Confirmation (measurement)
0010Confirmation (deploy)
0011Confirmation (pre-deploy)
0012Data retrieval (internal processing in progress)
DEPTIME
0 – DEPLOY command has not been received.
< 0 – Number of seconds until deployment starts.
> 0 – Number of seconds that deployment has been running.
MEASTIME
0 – START command has not been received.
> 0 – Number of seconds that measurement has been running.
CURRTIME
The current instrument time
Time format is “YYYY-MM-DD HH:MM:SS
WAKEUP
Reason for instrument wakeup
0 – Bad power
1 – Power on
2 – Break
3 – Real-time clock
INTPROC
Internal processing Active

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nortek Signature Series and is the answer not in the manual?

Nortek Signature Series Specifications

General IconGeneral
TypeAcoustic Doppler Current Profiler (ADCP)
Operating Temperature-5°C to +45°C
Storage Temperature-20°C to +60°C
Temperature SensorIntegrated
Pressure SensorIntegrated
Compass and Tilt SensorIntegrated
Data StorageInternal memory
Communication InterfacesEthernet, RS-232
Depth RatingUp to 6000 meters (depending on model)
Data OutputASCII, binary
Beam AngleVaries by model
Velocity Accuracy±0.5% of measured velocity ±1 mm/s