A-38
StatStrip Glucose Hospital Meter
Table A-56 Study #2 and #3 (Retrospective Combined)
Fingertip capillary samples with glucose concentrations
<75 mg/dL
Within
+5 mg/dL
Within
+10 mg/dL
Within
+12 mg/dL
Within
+15 mg/dL
Exceeds
+15 mg/dL
907/1894
(47.9%)
1470/1894
(77.6%)
1614/1894
(85.2%)
1737/1894
(91.7%)
157/1894
(8.3%)
Fingertip capillary samples with glucose concentrations
>75 mg/dL
Within
+5%
Within
+10%
Within
+12%
Within
+15%
Within
+20%
Exceeds
+20%
7473/
14884
(50.2%)
11807/
14884
(79.3%)
12799/
14884
(86.0%)
13712/
14884
(92.1%)
14350/
14884
(96.4%)
534/
14884
(3.6%)
A.9 Wireless Meter Use
The StatStrip Glucose Hospital Meter with wireless option
is an option that is factory installed before purchase.
A.9.1
Wired & Wireless Network Unauthorized Access Protection
When the StatStrip Meter is connected to a local area wire-
less 802.11a/b/g network, these networks must be properly
secured. It is the responsibility of the System Administrator
and Local Network Support to:
• Take all reasonable measures, both physically and
electronically to protect the local area network and
wireless networks on which the meters are con-
nected.
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 38 9/13/18 8:29 AM