1-15
1. Intro.
Instructions for Use Manual 1 Introduction
1.9.3 Meter Sleep/Wake Up
The LCD display is turned off to conserve battery power
(sleepmode)afteranOperatordenedtimeofnoactivity.
Keep-awake activities include:
• Touching the screen
• Placing the meter into the Charging Station
• Inserting a test strip
If the meter is placed into the docking station, the following
conditions should be expected:
• If Patient Result is the currently displayed screen
when docking occurs, the results are auto-saved.
• If the currently displayed screen is a Setup screen,
anyunconrmedinputdataormenuselectionis
discarded upon docking.
Wake Up
When in the sleep mode, the following conditions activate
the meter: the meter displays the last screen it displayed
before it went to sleep. To wake the meter, one of the fol-
lowing can be done:
• Touching the screen
• Inserting a strip (test/QC/Linearity)
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 15 9/13/18 8:29 AM