StatStrip Glucose Hospital Meter
1.9.4 Result Alerts
The result is displayed differently depending on whether it
is in or out of the normal range for glucose measurement.
Results within the normal range are displayed in
Blue.
Results outside the normal range are displayed in
Red.
If the value is outside the technical range of the me-
ter, the low or high end of the technical range value
displays as <XX or >YY (where XX-YY represents
the technical range).
A Single up arrow () is displayed for a result if the
value is higher than the upper end of the normal
range but within the critical range.
A double up arrow (↑↑) is displayed for a result if
the value is higher than the upper end of the critical
range.
A Single down arrow () is displayed for a result if
the value is lower than the lower end of the normal
range but within the critical range.
A double down arrow (↓↓) is displayed for a result
if the value is lower than the lower end of the critical
range.
Figure 1.3 Patient Result: Exceeds Measure/HI screen
13:42
Pt ID: 194546646
Meltzer, Jessica
Strip Lot: 0206806099
OK
Reject
Accept
Comment
Glu
HI
mg/dL
Exceeds Measure >600
Op: ****
Patient
07/19 06:19
Pending
Normal 70-130
Name: A Unit-1
HI Displayed
Result >600
Double
Up Arrows
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 16 9/13/18 8:29 AM