A.9.8 WiFi Function and Performance
As described above, the StatStrip Glucose Hospital Meter
with wireless option supports the 802.11a, b & g standards.
In a typical WLAN environment (correct WLAN administra-
tion, “good” signal quality, typical population of other clients
present, any of the supported security models enabled, etc),
thistranslatesintothefollowingWiFispecicperformance:
All communication between the StatStrip Glucose
Hospital Meter with wireless option and a suitable
device manager is in the POCT1-A communication
standard.
The StatStrip Glucose Hospital Meter with wireless
option is capable of downloading from a suitable
devicemanger,viaWLAN,acompletecongura-
tion set, including up to 1000 Patient records, 4000
Operator records and 30 reagent lot numbers, when
operated in a typical WLAN environment.
The
StatStrip Glucose Hospital Meter with wire-
less option
r, with WiFi enabled and connected, will
immediately transfers Patient and QC Results to a
suitable device manager at the completion of the
test and result entry into the meter database. Upon
receipt of the result, the device manager replies to
the meter with an “ACK” indicating the result was
properly received. The meter then marks the result
as “Transmitted” in its database. If the device man-
ager does not reply with the appropriate “ACK,” the
meter will not mark the result in the database, and
attempt to resend the result the next time the meter
communicates with the device manager. Complete
result download and acknowledgement cycle in a
typical WLAN environment is less than 1 second.