A-46
StatStrip Glucose Hospital Meter
Tochangeanyofthemetercongurationswirelessly,in-
cluding the radio settings and/or meter setup, the StatStrip
Glucose Hospital Meter with wireless option must be placed
in the docking station with the radio turned on and properly
congured.
A.9.7 WiFi Technical Considerations
Prior to connecting any StatStrip Glucose Hospital Meter
with wireless option to a wireless network, Nova Biomedical
recommends that a WLAN site survey be performed.
The WLAN site survey
• To ensure that Wireless Access Points will provide
enough coverage
• The performance is enough to support the StatStrip
Glucose Hospital Meter with wireless option.
• Detail RF signals, including all existing WLANs
along with any competing RF signals
• Interferences (building structure related and other
wireless equipment / devices).
Recommendation of a Wireless Meter implementation:
• A minimum of one wired docking station be con-
nected via an RJ-45 network jack per usage area.
• A networked docking station provides redundancy
if a wireless network malfunctions, loses service,
or if the meter radio fails to operate correctly.
• If used in an area with low signal or interferences,
it is recommended to install a connected wired
docking station for redundancy.
• The connected docking station allows immediate
bidirectional transmission of data when the radio
is turned off and the meter is docked.
See A.8.8 for complete StatStrip Glucose Hospital Meter with
wirelessoptionradiospecications,availableconguration
parameters, and security and encryption options.
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 46 9/13/18 8:29 AM