1-8
StatStrip Glucose Hospital Meter
1.8 Interfering Substances
Glucose Interferences:
The StatStrip Glucose Hospital Meter System exhibits no
interference from the following list of drugs or physiological
conditions commonly found in patients in intensive care
settings with complex medical conditions and drug regiments
up to the following concentration levels:
Tested
Interfering Substances
Tested
Concentration Level
Acetaminophen 20.0 mg/dL 1.323 mmol/L
Acetoacetate 51.0 mg/dL 5 mmol/L
Acetone 69.7 mg/dL 12 mmol/L
Acyclovir 0.6 mg/dL 0.027 mmol/L
Albuterol 0.06 mg/dL 0.0025 mmol/L
Amitriptyline 0.06 mg/dL 0.0022 mmol/L
Amoxicillin 5 mg/dL 0.137 mmol/L
Ampicillin 0.8 mg/dL 0.0229 mmol/L
Ascorbic Acid 22.5 mg/dL 1.278 mmol/L
Atropine 0.01 mg/dL 0.000346 mmol/L
Beta-hydroxybutyrate 166.6 mg/dL 16 mmol/L
Bilirubin 29.3 mg/dL 0.5 mmol/L
Captopril 0.6 mg/dL 0.0276 mmol/L
Carbamazepine 1 mg/dL 0.0423 mmol/L
Cefaclor 35 mg/dL 0.952 mmol/L
Cholesterol 1000 mg/dL 25.86 mmol/L
Cimetidine 5 mg/dL 0.198 mmol/L
Citric Acid 384.3 mg/dL 20 mmol/L
Creatinine 6 mg/dL 0.53 mmol/L
Digoxin 0.3 mg/dL 0.0038 mmol/L
Diltiazem 0.1 mg/dL 0.0024 mmol/L
Dopamine 20 mg/dL 1.306 mmol/L
Enalapril 0.5 mg/dL 0.0133 mmol/L
Ephedrine 1 mg/dL 0.0605 mmol/L
Ethanol 399.9 mg/dL 86.8 mmol/L
Famotidine 0.042 mg/dL 0.0012 mmol/L
Fluconazole 2 mg/dL 0.0653 mmol/L
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 8 9/13/18 8:29 AM