2-2
StatStrip Glucose Hospital Meter
• Whenever problems (storage, operator, instrument)
are identied or anytime there is a concern the
accuracy of the meter may have been affected by
rough handling (such as dropping the meter).
• As required by the institution’s quality control policy
or local regulatory requirements
Good Laboratory Practice principles suggest that external
controls must be run whenever the laboratory director has any
question about the test system integrity or operator technique.
2.3 Quality Control Test
The following section explains how to run a Quality Control
Test with one of the three StatStrip Glucose Control Solutions.
1. From the Patient Test screen, press the QC soft
key.
2. The Enter Strip Lot screen displays. Enter the Strip
Lot Number or scan the barcode. To scan the bar-
code, press the Scan soft key.
NOTE: If the Strip Lot Number is invalid, the screen displays
the invalid number with "is not a valid Strip Lot # Try
again."
3. Press the Accept soft key if the lot number is correct.
Figure 2.1 Enter Strip Lot Screens