A-48
StatStrip Glucose Hospital Meter
•
The StatStrip Glucose Hospital Meter with wireless op-
tion, with WiFi enabled and connected, will immediately
query the device manager if an unknown Operator or
Patient ID is entered on the meter during a test proce-
dure. If either the Operator or Patient ID is not valid for
that meter at the time of the query, the device manager
will reply to the meter with corresponding information
for the subject. If the ID is valid, the meter will add
the data to its operator or patient list in its database.
Latency time of up to 10 seconds for query reply is a
denablecomponentofthemetercongurationset.
Average range for direct connection between the StatStrip
Glucose Hospital Meter with wireless option (WiFi on) and
the WAP:
• 49 to 66 feet (15 to 20 meters) with air, direct view,
and low disturbances
•
The actual range depends on the positioning of the
WAP’s antennas and other topological properties of
the space between StatStrip Glucose Hospital Meter
with wireless option with WiFi enabled and WAP.
• Also, dynamic control of the transmitting power of
the access point may reduce the maximum distance
between StatStrip Glucose Hospital Meter with
wireless option and WAP where communication
can be guaranteed.
The StatStrip Glucose Hospital Meter with wireless option
with WiFi enabled:
• Designed to coexist in environments with other
wirelessly communicating devices
•
Does NOT provide any real-time wireless functionality
for the transmission, receipt, or monitoring of data
• Communication is described as exclusive, single,
digitaldataelds,anddoesNOTcommunicatein
a continuous waveform.
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 48 9/13/18 8:29 AM