6-5
6. Maint.
Instructions for Use Manual 6 Maintenance
1. Clean the Meter.
Remove a fresh germicidal wipe from the canister.
Wipe the external surface of the meter thoroughly
with a fresh germicidal disinfecting bleach wipe.
Discard the used wipe into an appropriate biohazard
container.
2. Disinfect the Meter.
Using a new, fresh germicidal bleach wipe,
thoroughly wipe the surface of the meter (top,
bottom, left, and right sides) a minimum of 3 times
horizontally followed by 3 times vertically avoiding
the bar code scanner and
electrical connector.
Gently wipe the surface area
of the test strip port making
surethatnouidentersthe
port.
Figure 6.6 Wiping the Surface of the Meter
3. Observe surface contact time.
Ensure the meter surface
stays wet for 1 minute and
is allowed to air dry for an
additional 1 minute.
Figure 6.7 Ensure Surface Stays Wet for a Minimum of 1 Minute
4. Dispose of wipe and gloves.
Dispose of used wipe and gloves
in a standard biohazard container.
Figure 6.8 Dispose of Wipe and Gloves
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 5 9/13/18 8:29 AM