APPENDIX C - SOFTWARE SERVICE TOOL INSTRUCTIONS
Ohaus Corporation www.ohaus.com 72 Explorer
®
Series Service Manual
TABLE C-1 INTERFACE COMMAND LIST (CONTINUED)
Set Percent application reference weight (x) in grams. (must have reference weight stored)
Print Percent application reference weight.
Set Dynamic Weigh Level to x. (x = 1 - 99 seconds)
Set Dynamic Weigh Mode. x = A (Automatic), S (Semi-Automatic), M (Manual)
Print Dynamic Weigh Level.
Start Dynamic Weigh cycle. (Manual Mode)
Clear locked weight (weight < threshold) in Dynamic Weigh, Display Hold & Totalize.
Set Checkweighing Over Limit in grams x.
Set Checkweighing Under Limit in grams x.
Print Checkweighing Over Limit.
Print Checkweighing Under Limit.
Set Checkweigh mode (1=over/under, 2=target/% tolerance, 3=Target/weight tolerance
Set Checkweighing target in grams x for percent tolerance mode.
Print Checkweighing Target for percent tolerance mode.
Set Checkweighing target in grams x for weight tolerance mode.
Print Checkweighing Target for weight tolerance mode.
Set Checkweighing % tolerance x. Attention: when x is a positive value, it is used to set the
+tolerance value; vice versa.
Print Checkweighing % tolerance.
Set Checkweighing weight tolerance x. Attention: when x is a positive value, it is used to set
the +tolerance value; vice versa.
Print Checkweighing weight tolerance.
Set Display Hold mode (Peak Hold). x = A (Automatic), S (Semi-Automatic), M (Manual)
Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON).
Set filter level to x (1 = low, 2 = med, 3 = high).
Set Auto-zero to x (x = 1 for 0d, x = 2 for 0.5d, x = 3 for 1d, x = 4 for 3d).
Resets all Balance menus to factory defaults. Attention: The binary code of this commands is
“1B20520D0A”or“1B520D0A”.
Program user Name. Attention: only allowed numeric input.
Set Totalize Mode. x = A (Automatic), M (Manual).
Set Time, x format: hh:mm:ss.
Set Date, x format: mm/dd/yyyy.
Continuous weight, same as CP.
Stable load, same as SLP.
Interval Print x = interval in sec (1-3600) 0 = off, same as xP.
Set AutoPrint off, same as 0P.
Begin Span Cal, same as C.
Set Animal Mode to Auto, Semi-Auto, Manual. Same as xAW(A/S/M)
Prints current mode, same as PM.
0 = print unstable data, same as IP; 1 = print stable only
1)
, same as SP.
0 = disableresponse1=enableresponse.Attention:thiscommandonlycontrolsthe“OK!”
response.