APPENDIX B - SERVICE MENU
Explorer Series Service Manual 69 Ohaus Corporation www.ohaus.com
INTERNAL REPEATABILITY TEST
The internal weight is used to determine the repeatability of the load-cell. The standard deviation is
determined after 10 cycles of the internal weight.
TABLE B-1 SERVICE CALIBRATION MASS VALUES
Balance Info
Contains information about the balance including:
Balance Type, Max, d, Balance ID, IDNR, Loadcell Type, Loadcell Serial Number, Terminal Serial
Number, Terminal SW Version, Base SW Version.
Exit Service Mode
Exits the Service Mode and returns to Normal Mode.