94
Chapter 6 Compatible Reprocessing Methods
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Suction
cleaning
adapter
(MH-856)
Cleaning
adapter for
instrument
channel port
(MAJ-350)
Washing tube
(MH-974)
Ultrasonic cleaning
*1
Manual
cleaning
Alkaline enzymatic detergent
Neutral enzymatic detergent
Manual
disinfection
Peracetic acid
Glutaraldehyde
Drying Alcohol
Automatic
cleaning
and
disinfection
AER ETD Double (Peracetic acid)
ETD4 (Peracetic acid)
ETD4 (Glutaraldehyde)
OER-AW
*2
(ACECIDE)
*3
WD (alkaline detergent, thermal disinfection)
Sterilization Hydrogen
perioxide
*4
V-PRO
®
maX
STERRAD
®
NX
®
with
ALLClear™ Technology
STERRAD
®
NX
®
STERRAD
®
100NX
®
with
ALLClear™ Technology
STERRAD
®
100NX
®
STERRAD
®
100S
Steam sterilization (autoclaving)
Ethylene oxide gas
Low Temperature steam and formaldehyde
(LTSF)
compatible not compatible
Table 6.5 List of compatible methods
*1 The endoscope is only compatible with ultrasonic cleaning as performed in an
Olympus-recommended reprocessor, such as OER-AW (OER-AW may not be available in some
areas). When using an AER/WD that is recommended by Olympus other than listed above,
contact Olympus.
*2 OER-AW is not available in the member states of the EU.
*3 ACECIDE disinfectant solution, which is peracetic acid, is exclusively for an
Olympus-recommended endoscope reprocessor, such as OER-AW (ACECIDE may not be
available in some areas).
*4 The cycles that are not listed in Table 6.1 through 6.5 can not be used for reprocessing.