Chapter 6 Compatible Reprocessing Methods
97
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Do not reuse rinse water.
Rinsing after disinfection
Olympus recommends using sterile water, fresh potable water or water that
has been processed (e.g., filtered, deionized, or purified) to improve its
chemical and/or microbiological quality.
Some national or professional guideline recommend using sterile water for
rinsing endoscopes, accessories, and medical instruments. Consult with
your institution’s infection control committee regarding local policies on
water quality.
6.7 Alcohol
• Do not reuse alcohol.
• Alcohol is not a sterilant or high-level disinfectant.
Use Medical-grade 70% ethyl or 70% isopropyl alcohol.