Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
9.2 Preparing the equipment for reprocessing
If the endoscope is not immediately precleaned after each
procedure, residual organic debris will begin to solidify and it
may be difficult to effectively reprocess the endoscope.
Preclean the endoscope at the bedside in the procedure room immediately after
each procedure. These steps are to be performed when the suction pump is still
connected to the endoscope. During precleaning, wear appropriate personal
protective equipment.
Equipment needed
Prepare the following equipment, and wear appropriate protective equipment.
9.3 Precleaning the endoscope and accessories
Preparation
1.
Turn the ultrasound center, video system center and light source OFF.
2. Prepare a 500 cm
3
(500 ml) container of detergent solution at the
temperature and concentration recommended by the detergent
manufacturer.
3. Prepare clean water in a 500 cm
3
(500 ml) container.
• Personal protective equipment
• Clean, lint-free cloth
•500cm
3
(500 ml) containers
• Detergent solution
• Clean water
• Air/water channel cleaning adapter (MAJ-629)
• Washing tube (MH-974)
•5cm
3
(5 ml) syringe