Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
9.5 Manually cleaning the endoscope and
accessories
To prevent a water leak, do not apply excessive force when
cleaning the endoscope.
After completing the leakage test, perform manual cleaning according to the
procedures described below.
In case of excessive bleeding and/or delayed reprocessing, perform “Presoak for
excessive bleeding and/or delayed reprocessing after each procedure” on
page 157 and follow the procedure described below.
Reusable parts that are normally reprocessed with the
endoscope
Equipment needed
Prepare the following equipment and wear appropriate personal protective
equipment.
• Channel plug (MAJ-621)
• Injection tube (MH-946)
• Water-resistant cap (2 pcs) (MH-553)
• Chain for water-resistance caps (MAJ-1739)
• Washing tube (MH-974)
• Cleaning adapter for instrument
channel port
(MAJ-350)
• Personal protective equipment
•30cm
3
(30 ml) syringe
•5cm
3
(5 ml) syringe
• Clean, lint-free cloths
• Large basins
• Detergent solution
• Clean water
• Suction pump
• Suction cleaning adapter (MH-856)
• Single use combination cleaning brush (BW-412T)
• Cleaning brush (MAJ-1534)
• Single use single-ended cleaning brush (BW-400L)
• Channel plug (MAJ-621)