Chapter 11 Reprocessing Endoscopes and Accessories Using an AER/WD
207
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
When setting the endoscopes in the ETD, put the forceps
elevator in intermediate position of the range of movement by
moving the elevator control lever and refer to the instruction
manual for the ETD.
When using the ETD, conduct all steps of precleaning and
manual cleaning as instructed in this manual before setting
the endoscope in the ETD.
(Optional) Alcohol flush
Refer to “(Optional) Alcohol flush” on page 155 and page 179 as the procedure
is same.
Dry the endoscope and accessories
Refer to “Dry the endoscope” on page 156 and “Dry the accessories” on
page 182 as the procedure is same.
11.3 Reprocessing endoscopes and accessories
using an OER-AW
OER-AW
If the endoscope has a forceps elevator, raise the forceps elevator to an
angle of approximately 45° and set it so that the back of the forceps
elevator can be sufficiently cleaned and disinfected.
Reprocess the endoscope using the OER-AW.
For the information concerning the details of the reprocessing steps
required , please refer to the instruction manual for the OER-AW before
putting this endoscope into the OER-AW.
Make sure to use the correct adapters for this endoscope according to
the instruction manua for the OER-AW or Section 6.9, “OER-AW
(Olympus Endoscope Reprocessor)”.
OER-AW is not available in the member states of the EU.