EasyManua.ls Logo

Olympus EVIS EXERA II - Signs of Degradation from Reprocessing and Its Number of Times

Olympus EVIS EXERA II
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
106
Chapter 6 Compatible Reprocessing Methods
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
6.13 Signs of degradation from reprocessing and its
number of times
When reprocessing devices with chemicals, AER and WD as described above,
you will eventually start to see signs of degradation. There are also limits to the
number of times a given reprocessing method can be used on a particular
device.
Improper reprocessing shown below may significantly reduce
the life time of a medical device.
Reprocessing without observing the instructions of the
manufacturer
Application of multiple sterilization methods
Immersion in a chemical for an excessive amount of time
Endoscope
Reprocessing may cause the following degradations. If any of these
signs of degradation happens, contact Olympus.
Cracked, peeled or discolored adhesives at either end of
the bending section and at objective lens and light guide
lens
Cracks in the bending section cover
Peeling or wrinkling of the insertion tube
Irregularity of the ultrasound image (e.g., excessively
dark, not clear, or not visible)
Figure 6.4
Deteriorated adhesive
exposing thread
Bending section cover
Cracks in adhesive
Distal end
Deteriorated missing
adhesive around lens
Cracks in bending
section cover
Peeling in
adhesive

Table of Contents

Other manuals for Olympus EVIS EXERA II

Related product manuals