EasyManua.ls Logo

Olympus EVIS EXERA II - Chapter 11 Reprocessing Endoscopes and Accessories Using an AER;WD; Reprocessing Endoscopes and Accessories Using an AER; Reprocessing Endoscopes and Accessories Using an ETD

Olympus EVIS EXERA II
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
206
Chapter 11 Reprocessing Endoscopes and Accessories Using an AER/WD
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Chapter 11 Reprocessing Endoscopes
and Accessories Using an
AER/WD
11.1 Reprocessing endoscopes and accessories
using an AER
Follow the workflow described in Section 8.2, “Workflow for reprocessing
endoscopes and accessories” when reprocessing endoscopes and accessories
with an AER.
Be sure to attach all required connectors to the endoscope and accessories. For
details concerning appropriate connectors, refer to the instructions of the AER
manufacturer.
Manually clean and disinfect any endoscopes and accessories that are not
compatible with the AER.
After reprocessing the endoscope using an AER, dry the
electrical contacts of the endoscope connector not by
air-drying but by wiping with sterile lint-free cloths. Otherwise,
hard water residue may be deposited on the electrical
contacts, which may cause an abnormal endoscopic image
when the endoscope is used.
11.2 Reprocessing endoscopes and accessories
using an ETD
ETD
Reprocess the endoscope using the standard cycle in the ETD.
For the information concerning the details of the reprocessing steps
required, please refer to the instruction manual for the ETD before
putting this endoscope into the ETD.
Make sure to use the correct adapters for this endoscope according to
the instruction manual for the ETD or Section 6.8, “ETD (Endo Thermo
Disinfectors)”.

Table of Contents

Other manuals for Olympus EVIS EXERA II

Related product manuals