Chapter 6 Compatible Reprocessing Methods
95
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Endo-therapy accessories which are marked by the words
“AUTOCLAVE” or “AUTOCLAVABLE”, or accessories with a
green model reference label are compatible with autoclaving.
6.3 Detergent solution for manual cleaning
• Excessive foaming prevents detergent solution from properly
contacting the surfaces and channel walls of the endoscope
and accessories, and may impair effective cleaning.
• Do not reuse detergent solution.
Use a neutral (20 – 45°C) or alkaline (20 – 40°C, <pH 10.8) low-foaming
enzymatic detergent with no abrasion labeled for use with flexible endoscopes
and accessories that has been approved by your national regulatory agency for
use in reprocessing flexible endoscopes, accessories, and medical instruments.
Follow the instructions provided by the detergent manufacturer regarding
concentration, temperature, contact time, use life, expiration date, and rinsing
unless otherwise specified by Olympus.
The detergents shown in Table 6.6 were used for validation.
Trade name Type Manufacturer
neodisher
®
Mediclean forte
Alkaline enzymatic Dr. Weigert
Endozime
®
AW
Neutral enzymatic Ruhof
Table 6.6 Detergents used for validation