100
Chapter 6 Compatible Reprocessing Methods
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
When the endoscope is cleaned and disinfected
simultaneously in combination with an endoscope of the
same or another model using OER-AW, see the group
number of the endoscopes shown in the “List of Compatible
Endoscopes/Connecting Tubes 2 and the instruction manual
for OER-AW to confirm the combination. The group number
of the endoscope model should be listed in Table 6.12.
6.10 Washer-Disinfector
Use only a WD whose compatibility for cleaning medical
instrument has been certified by the WD’s manufacturers,
and which meets ISO 15883-1, ISO 15883-2 requirement.
Using an incompatible WD may pose an infection control risk
and may considerably damage the medical instrument.
Perform manual cleaning of the medical instrument before
loading them into a WD as described in the instruction
manual. If the medical instrument are not properly cleaned,
subsequent reprocessing will be ineffective.
For details on the operation of the WD, refer to the instruction
manual of the WD.
Do not use an undesignated detergent, neutralizer, or WD
program. This could damage the medical instrument.
Follow the instructions provided by the process chemical
manufacturer regarding concentrations, temperature and
exposure time of detergent and neutralizer.
Regularly maintain the WD as recommended by its
manufacturer.
This endoscope cannot be loaded in the WD. For details, see
Table 6.1 through 6.5.
*1 These products are not available in the member states of the EU.
OER-AW
GF-UCT180 Group 3
Table 6.12