Chapter 3 Preparation and Inspection
53
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Inspection of the ultrasound image with the Hitachi or
FUJIFILM Healthcare diagnostic ultrasound system
1.
Inspect the diagnostic ultrasound system as described in its instruction
manual.
2. Turn the diagnostic ultrasound system ON.
3. Confirm that the ultrasound image is visible on the diagnostic ultrasound
system (see Figure 3.24).
Figure 3.24
4. Press the “FREEZE” switch to change the ultrasound image to the FREEZE
mode.
3.7 Preparation and inspection of the balloon
• Balloons used with this instrument contain natural rubber
latex that may cause allergic reactions in some individuals.
Do not use the balloon on a latex-sensitive patient. Instead,
perform the procedure using “The sterile deaerated water
immersion method” described in Section 4.2, “Observation of
the ultrasound image” on page 66.
• When using the sterile balloon (MAJ-249), do not use an
instrument after the expiration date displayed on the sterile
package. Doing so may pose an infection control risk.
• The balloons are disposable, and are intended for a
single-use only. A new one must be used for each patient. Do
not attempt to reuse the balloon. This could pose an
infection-control risk or cause equipment damage.
Ultrasound image (B-mode)