54
Chapter 3 Preparation and Inspection
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Confirm that the balloon applicator (MAJ-675) and cotton
thread that is used to tie the balloon have been properly
reprocessed as described in Chapters 5, “Reprocessing:
General Policy” through 7, “Cleaning, Disinfection, and
Sterilization Procedures”.
If you find that a sterile balloon’s peel pack is open,
damaged, or soiled, do not use that balloon.
When using the sterile balloon (MAJ-249), inspect the sterile
package for tears, inadequate sealing, or water damage. If
the sterile package shows any irregularity, the sterile
condition of the instrument may have been compromised.
The balloons and the balloon applicator to be used with this
endoscope are listed below. Balloon 3 is shipped sterile.
The balloon can easily tear, so beware of sharp objects.
Store the balloon in a cool environment with low humidity.
The resealable package containing the balloons also
contains a deoxidizer to maintain a deoxygenated condition
until the sterile peel pack is opened. After the resealable
package is opened, the balloons will gradually deteriorate. To
minimize deterioration, always keep the package sealed.
The balloon applicator is reusable and supplied nonsterile.
Prepare the balloon and the balloon applicator.
Balloon 3 (sterile) MAJ-249
Balloon applicator MAJ-675