Chapter 4 Operation
79
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
4.6 Transportation of the endoscope
Transporting within the healthcare facility
When carrying the endoscope by hand, loop the universal cord, hold the
endoscope connector with the control section in one hand and hold the distal
end of the insertion tube securely, but gently without squeezing, in the other
hand (see Figure 4.7).
Figure 4.7
Transporting outside the healthcare facility
Transport the endoscope in the carrying case.
Always clean, disinfect, or sterilize the endoscope after
removing it from the carrying case. If the endoscope is not
cleaned, disinfected, or sterilized, it could pose an infection
control risk.
The carrying case cannot be cleaned, disinfected, or
sterilized. Clean and disinfect or sterilize the endoscope
before placing it in the carrying case.
To avoid damage to the endoscope caused by changes in air
pressure, do not attach the water-resistant cap when
transporting the endoscope.
Before putting the endoscope in the carrying case, always
make sure that the forceps elevator is not raised. Putting the
endoscope in the carrying case while the forceps elevator is
raised could damage the endoscope.

Table of Contents

Other manuals for Olympus EVIS EXERA II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus EVIS EXERA II and is the answer not in the manual?

Olympus EVIS EXERA II Specifications

General IconGeneral
BrandOlympus
ModelEVIS EXERA II
CategoryMedical Equipment
LanguageEnglish

Related product manuals