Chapter 3 Preparation and Inspection
41
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
3.5 Inspection and connection of ancillary
equipment
Inspection of ancillary equipment
• Attach the water container to the specified receptacle on the
trolley (cart) or the light source. If the water container is
attached anywhere else, water may drip from the water
container’s water supply tube, and equipment malfunction
can result.
• Take care not to spill water from the water container’s
connection adapter when detaching the connection adapter
from the endoscope. Spilled water could splash on the
equipment, and may cause equipment malfunction.
1. Prepare and inspect the light source, video system center, ultrasound
diagnostic equipment, monitor, water container, suction pump, and
EndoTherapy accessories as described in their respective instruction
manuals.
2. Confirm that there are no scratches, cracks, excessive wear, or deformation
of the ultrasonic cable.
Connection of the ultrasonic cable (MAJ-1597) and Olympus
universal endoscopic ultrasound center
Insert the ultrasonic connector properly into the transducer port of the Olympus
universal endoscopic ultrasound center. Rotate the connector’s lock handle 1/4
turn clockwise (see Figure 3.15).
Figure 3.15