Chapter 3 Preparation and Inspection
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Connection of the ultrasonic cable (MAJ-1597) and Hitachi or
FUJIFILM Healthcare diagnostic ultrasound system
Insert the ultrasound connector properly into the transducer port of the
diagnostic ultrasound system. Rotate the connector’s lock handle 1/4 turn
clockwise (see Figure 3.16).
Connect the Ultrasound connector to one of the probe
connectors on the lower side (PROBE 3 or 4) of the
ProSound F75 (see Figure 3.16).
If the Ultrasound connector is connected to one of the probe
connectors on the upper side (PROBE 1 or 2) , the operation
panel of the ProSound F75 may hit the Ultrasound connector,
which may result in the equipment damage. When adjusting
the height, a horizontal and / or vertical position of the
operation panel of the ProSound F75 with the Ultrasound
connector connected to the PROBE 1 or 2, move slowly the
operation panel with visually confirming the position of the
bottom of the operation panel.
Lock handle
Ultrasound connector
Transducer port
SSD-α10