Chapter 5 Reprocessing: General Policy
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
When aerating or irrigating the endoscope channels, the air
or water pressure must not exceed 0.5 MPa (5 kgf/cm
2
,
72 psig). Higher pressures may cause damage to the
endoscope.
When reprocessing an endoscope, confirm that the two
water-resistant caps are securely attached to the endoscope
connector before immersion in reprocessing fluids. If two
water-resistant caps are not securely attached, water,
detergent solution and/or disinfectant solution could enter the
endoscope and damage the equipment.
5.4 Reprocessing before the first use
New endoscopes, repaired endoscopes, accessories, and the carrying case for
endoscopes are not reprocessed prior to shipping from Olympus, regardless of
whether those instruments are for new purchase, demo, or loaner purposes.
Reprocess all such endoscopes and accessories received from Olympus
according to the instructions given in this manual before storage and before
using them in a patient procedure.
5.5 Reprocessing and storage after use
Do not reuse detergent solution.
Do not reuse rinse water.
Detergent and disinfectant solutions are only effective when
used according to the detergent and disinfectant
manufacturer’s instructions. Follow the manufacturer’s
instructions regarding activation (if required), concentration,
temperature, contact time, use life, and expiration date
required to achieve successful cleaning and disinfection.
If the disinfectant solution is reused, check its efficacy by
proper methods, such as using a test strip, according to the
disinfectant manufacturer’s recommendations prior to use.
Do not reuse alcohol.
Alcohol is not a sterilant or high-level disinfectant.
To maintain sterility of equipment following sterilization, use
sterilization pouches and wraps according to national
guidelines.