102
Chapter 6 Compatible Reprocessing Methods
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
ISO 15883-2 regulate that disinfection temperature give an
A
0
of at least 600. Olympus recommends that WD provide an
A
0
value of 3000.
Do not use a program that exceeds 55°C in the cleaning
process and 93°C in the disinfection process.
Do not use a program that exceeds 10 minutes in the
disinfection process.
Confirm that there are no sudden temperature fluctuations in
the program you have selected.
Olympus performed validation using a WD compliant with
ISO 15883-1, 2 requirements.
Detergent solution for WD
Use a low-foaming, medical-grade, alkaline enzymatic detergent (20 –
55°C, <pH 10.8).
Use a detergent whose compatibility for cleaning medical instrument is
certified by the WD’s manufacturer, and which are approved by a
competent authority. Follow the instructions provided by the detergent
manufacturer.
The detergent shown in Table 6.14 was used for validation.
Neutralizer
If rinsing with neutralizers after cleaning/disinfection is
necessary, make sure to remove all traces of neutralizers by
using the water as described in the instruction manuals for
WD.
Use neutralizers recommended in the instruction manuals for the WD
and detergent.
Follow the instructions provided by the neutralizer manufacturer
regarding concentration, temperature, contact time and expiration date.
Trade name Type Manufacturer
neodisher
®
Mediclean forte
Alkaline enzymatic Dr. Weigert
Table 6.14 Detergent used for validation