Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
113
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Reprocessing equipment parts and functions
For inspection of other equipment than that mentioned below, refer to the
instruction manual for the equipment being used.
Water-resistant cap (MH-553)
The two water-resistant caps are attached to the videoscope cable connector
and the ultrasonic cable connector on the endoscope to protect the connectors
from water penetration during reprocessing. For leakage testing, the venting
connector on the water-resistant cap must be connected to the leakage tester
(MB-155) (see Figure 7.2).
Figure 7.2
Chain for water-resistance caps (MAJ-1739)
This instrument has been designed to connect the water-resistant caps
(MH-553) to the endoscope. Please ensure that you do not forget to attach
the water-resistant cap prior to submersing the endoscope. This instrument
must be used to connect the water-resistant cap to a endoscope that has
either the S-cord connector mount or the chain connector (see Figure 7.3).
Figure 7.3
Groove
Seal
Venting connector
Notch
Chain part
Fitting part
Connecting plate
Hole